MS400 Multiparameter Simulator is a portable device, features in small volume and light weight. It can simulate accurately 12-lead ECG, Respiration, Temperature, 2-channel IBP or 4-channel IBP. And it can be tested in where the monitor is used.
Features
-
Simulate 35 kinds of arrhythmia (including Atrial, Ventricular, Premature Beat and Conduction Defects) and pacemaker waveforms (including Asynchronous, Demand Occ Sinus, Demand Freq Sinus, A/V Sequential, Non Capture, Non Function).
- Simulate four-channel IBP data.
- Simulate respiration signal with adjustable respiration rate, impedance and baseline impedance, and three kinds of apnea data.
- Simulate four kinds of temperature data.
- Simulate square wave, triangle wave, pulse wave and sine wave with adjustable frequency.
- Simulate standard ECG calibration, adult and pediatric ECG signal.
- ECG waveform complied with the standard of ANSI/AAMI EC13-2002, AHA, MIT ECG database, other ECG database and user-defined waveform data stored in TF card can be displayed by the simulator and exported to the equipments to be tested.
- Each waveform can be displayed by sliding the 3.5'' TFT touch screen, easy to operate.
- Power supply: rechargeable lithium battery or linear adapter, which reduces interference.
- Optional operation modes: keys and touch screen.
Parameters:
-
ECG
Conventional: All leads have the same voltage with the II-lead as the reference
Amplitude: 0.5,1.0,1.5 and 2.0mV
ECG Rate: 15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340 and 350 BPM
ECG artifacts: 50 and 60 Hz, muscle, baseline and breathing
The elevation of the ST segment: +0.8 to -0.8 mV, the step is 0.1 mV
-
Arrhythmia
Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia, Missed Beat, Atrial Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
Ventricular: Vent Fib (coarse), Vent Fib (fine), PVCs 6/min, PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5 PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
Premature Beat: Atrial PAC, Nodal PNC, PVC1 LV Focus, PVC1 E LV Focus, R on T LV Focus, PVC2 RV Focus, PVC2 E RV Focus, R on T RV Focus, Multifocal PVC
Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block
-
Four IBP Channels
The default: 00/00mmHg
Dynamic IBP(unit: mmHg)
Channel 1:
Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Ventricle: 25/00
Channel 2:Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
Channel 3:Artery: 120/80
Radial: 120/80
Left Ventricle: 120/00
Right Atrium(CVP): 15/10
Right Ventricle: 25/00
Pulmonary Artery: 25/10
Pulmonary Wedge: 10/2
Left Atrium: 14/4
Channel 4:
Auto Swan-Ganz (every 15 seconds): RA(CVP), RV, PA, PAW (Pulmonary Wedge)
Man Swan-Ganz: press ENTER key to change every time
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg
Channel 3: 0, 20, 40, 60, 80 and 100mmHg
Channel 4: P4 tracks the static pressure levels of P3
-
Respiration
Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIN
Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
Baseline impedance: 500, 1000, 1500, 2000Ω
Apnea: 12 s, 22 s, 32 s
-
Temperature
Four temperature: 0℃(32°F), 24℃( 75°F), 37℃(98.6°F), 40℃(104°F)
Accessories
Standard accessories 9V/1A(CE) power adapter, an User manual, a packaging box, a plastic packaging.