ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW
ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW
ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW
ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW
ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW
  • SKU:
  • Availability: in stock Many in stock Out of stock You can purchase this product but it's out of stock

ECG1201G Digital 12 Channel 12 Lead Electrocardiograph Thermal Printer PC SW

$799.00 $1,099.00
  • Barcode:
DESCRIPTION
Introduction
The device is an electrocardiograph which can collect 12-lead ECG signal simultaneously and print ECG waveform with thermal printing system. It features in, recording and displaying ECG waveform in auto / manual mode; measuring, analyzing and diagnosing ECG waveform parameters automatically; prompting for “Lead off” and “Lack of paper”; optional interface language (Chinese / English, etc.); built-in lithium battery, AC / DC; selecting rhythm lead freely to observe abnormal heart rhythm; case database management.

Functions
1)Adopt 7'' high-resolution color LCD, operate by buttons, convenient to use.
2)Sync collection for 12-lead ECG, support 12-lead and Cabrera-lead waveform display. Get high-quality ECG waveform via power frequency filter (50 / 60 Hz), baseline filter and EMG filter (25 / 35 Hz) of ECG signal.
3)Display 1 / 3 / 6 / 12 - lead ECG waveform, HR value, print mode, sensitivity, paper speed, filter state, battery power, background grid, etc. on one screen, and with prompt functions for lead - off, overload and system working state.
4)Power supply: AC / DC (which can adapt AC 50 / 60Hz), with built-in rechargeable lithium battery, charging circuit, perfect over-current and over-voltage protection circuit.
5)In optimal DC state, up to 10-hour standby time, continuous print for more than 3-hour, which meets the requirements of visiting a patient at home and body examination.
6)Built-in thermal printer, support multiple printing modes, such as, auto M × N, M × N + 1, M × N + 2, rhythm M line and manual, etc. The contents including time, paper speed, sensitivity, calibration signal, lead name, filter state and patient information can be printed. The waveform length printed, measurement parameter output, diagnosis conclusion, superimposed QRS waveform, histogram, trend chart and interval list, etc. can be set. With the timing print functions, which meet different requirements.
7)With the functions of auto-measurement and auto-interpretation for routine ECG parameters, provide measured results and auto-diagnosis conclusions for HR, PR interval, P Duration, QRS Duration, QT / QTc interval, P / QRS / T Axis, R (V5), S (V1), R (V5) + S (V1) amplitude, etc. which reduces the doctor’s burden.
8)Historical cases can be reviewed, inquired, modified, transmitted, printed, lead corrected, the cases can also be exported to other electronic file formats (dat, xml, bmp, png)
9)Multi-language (Chinese, English, Turkish, Portuguese, German, Russian and Kazakh, etc.) interface and report.
10)Support external USB standard keyboard, scanner and printer.

Performance
1)Input mode: floating and defibrillation protection
2)Lead: standard 12-lead
3)Patient leak current: < 10 µA
4)Input impedance: ≥ 2.5 MΩ
5)Frequency response: 0.01Hz-300Hz (-3dB)
6)Time constant: ≥ 3.2 s
7)CMRR: > 105 dB
8)Filter: power frequency (AC 50 / 60 Hz), EMG (25 Hz / 35 Hz (-3 dB)) and baseline drift
9)Recording mode: thermal printing system
10)Paper size: 210 mm (W) × 20 m (L), high-speed thermal paper
11)Paper speed:5 mm/s, 6.25 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s, error: ± 5 %
12)Gain (sensitivity): 2.5,5, 10, 20 ,40mm / mV, accuracy: ± 2 %
Standard sensitivity: 10 mm/mV ± 0.2 mm/mV
13)Record mode: auto, rhythm and manual record
14)Measurement parameters: HR, P - R interval, P Duration, QRS Duration, T Duration, Q - T interval, Q - Tc, P Axis, QRS Axis, T Axis, R (V5) amplitude, S (V1) amplitude, R (V5) + S (V1) amplitude
15)Safety class: class I, defibrillation-proof type CF applied part
16)Polarizing voltage: ± 610 mV
17)Noise level: ≤ 12 µVp-p
18)Sampling frequency for ECG signal input: 32 kHz
19)Sampling frequency for waveform data processing: 1 kHz
20)Sampling accuracy: 24 - bit
21)Minimum detection signal: a 10 Hz, 20 µV (peak - peak value) sinusoidal signal can yield a visual recorded deflection.
22)Pacemaker detection channel: all leads
23)Accuracy of input signal reproduction: ± 5 %
24)Amplitude quantization: ≤ 5µV/LSB
25)Time deviation in each channel: <100 µs
26)Power supply: 
AC: 100V ~ 240V (50 / 60 Hz)
DC: 14.8 V / 5200 mAh rechargeable lithium battery
27)Degree of protection against ingress of liquid: IPX0
28)Working mode: continuous operation

Accessories
One lead cable
One limb electrode
One chest electrode
One thermal recording paper
One power cord
One earth wire
One User Manual 

Physical characteristic
Operating:
a) Environment temperature: 5 ℃ ~ 40 ℃
b) Relative humidity: 25 % ~ 95 % (non - condensing)
c) Atmospheric pressure: 700 hPa ~ 1060 hPa
Transport and storage:
a) Environment temperature: - 20 ℃ ~ + 55 ℃
b) Relative humidity: ≤ 95 %
c) Atmospheric pressure: 500 hPa ~ 1060 hPa
Dimension: 340 mm (L) × 320 mm (W) × 86 mm (H)
Weight: 5 kg

Optional configuration
See BOM diagram.

ADDITIONAL INFORMATION
Color: Blue, Purple, White
Size: 20, 24
Material: 100% Polyester
REVIEWS
GENERAL TAB

RELATED PRODUCTS

BACK TO TOP