Introduction
The monitor has abundant functions that can be used for clinical monitoring with adult, pediatric and neonate. Users may select different parameter configuration according to different requirements. The monitor, power supplied by 100-240V~,50/60Hz, adopts 8'' color TFT LCD displaying real-time date and waveform. It can synchronously display eight-channel waveform and full monitoring parameters equipped with an optional external 48mm thermal recorder. The monitor can be connected to the central monitoring system via wire or wireless network to form a network monitoring system.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
Features
1)8'' TFT color LCD, multi-language interface (Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakh, Polish, Czech).
2)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
3)All-round monitor for adult, pediatric and neonate.
4)With standard interface, oxygen graph, trend graph, big character interface and view bed, convenient to observe.
5)Waveform, parameter color and location can be set optionally.
6)Finish all operations by keys and knobs.
7)Display 7-lead ECG waveform on one screen, cascade ECG waveform display
8)Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
9)Heart rate variability (HRV) analysis function
10)NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
11)Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
12)Drug concentration calculation and titration table functions.
13)One-touch printing of trend graph
14)Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
15)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
16)Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Performance
▲ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, 2-channel
3-lead, 1-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350 bpm
Accuracy: ±1 % or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2 mV ~ +2 mV
Accuracy: -0.8 mV~+0.8 mV ±0.04 mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
▲RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 12.5 mm/s, 25 mm/s
▲NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
▲PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
▲TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1℃
Accuracy: ±0.1℃
▲Power supply: 100-240V~, 50/60Hz
▲Safety classification: Class I, type CF defibrillation-proof applied part
Accessories
1)SpO2 probe extension cable
2)Adult fingertip SpO2 probe
3)Adult NIBP cuff
4)NIBP extension tube
5)ECG lead cable
6)ECG electrode
7)Temperature probe
8)Power cord
9)User Manual (GB only)on request for other Language
Physical characteristic
Dimension: 227mm(L)x153mm(W)x247mm(H)
Weight: 2.4 Kg