CA10M is designed for patients who can't breathe spontaneously, and needs to be used with a ventilator or anesthesia machine. It can't measure CO2 alone.
Introduction
Capnograph adopts mainstream method to measure the end-tidal carbon dioxide, airway respiration rate and other parameters. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow condition, as well as indirectly reflect the alveolar ventilation, which is applicable for use in pet hospital, emergency department, ICU, operating room, respiratory department and other occasions. The application population is animals. The Capnograph connects with patientâ respiratory tract by the airway adapter and breathing tube.
Features
1)End-tidal CO2 measurement function
2)Respiration rate measurement function
3)Over-limit alarm function for End-tidal CO2 and respiration rate
4)Low battery alarm function
Performance
End-tidal CO2 measurement:
1ïŒMeasurement range: 0 ~150mmHg
2ïŒResolution: 1mmHg
3ïŒAccuracy:
0ïœ40 mmHgïŒÂ±2 mmHg
41ïœ70 mmHgïŒÂ±5% of readings
71ïœ100 mmHgïŒÂ±8% of readings
101ïœ150 mmHgïŒÂ±10% of readings
Respiration rate measurement:
1ïŒMeasurement range: 2rpmïœ150rpm
2ïŒResolution: 1rpm
3ïŒAccuracy: ±1rpm
Safety characteristic:
1ïŒDevice type: internally powered equipment
2ïŒDegree of protection against electric shock: type BF applied part
3ïŒDegree of protection against ingress of liquid: IP22
Accessories
User Manual
CA10M Airway Adapter
Physical Characteristic
Dimension: 59Ă56Ă59.5mm
Weight: About 92g (with batteries)
Working environment
Temperature: 5â~40â
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 3V (2 AAA dry batteries)
Input power: â€1VA