CMS8500 Patient Monitor
The monitor boasts a plethora of functions suitable for clinical monitoring of adults, pediatrics, and neonates. Users have the ability to select parameter configurations tailored to specific requirements. Powered by 100V-240V~, 50Hz/60Hz, the monitor is equipped with a 14" color TFT LCD display presenting real-time data and waveforms. The device can simultaneously display ten-channel waveforms and complete monitoring parameters while offering an optional 48mm thermal recorder. Additionally, the monitor can be wirelessly or wiredly connected to a central monitoring system to form a network monitoring system.
This apparatus can monitor vital parameters, including ECG, RESP, NIBP, SpO₂, and dual-channel TEMP, among others. The parameter measurement module, display, and recorder are integrated into a single, compact, and portable device. Furthermore, its replaceable internal battery brings a great deal of convenience for patient mobility.
Function:
-
14" high-definition touch screen design, easy and convenient to operate, elegant and concise in appearance.
-
Independent physiological and technical alarm function, convenient for medical staff to quickly know patient's condition.
-
Adopt high-performance processor, stable, reliable and fast to process information.
-
Patient information can be rapidly entered, patient type (adult/neonate/pediatric) can be quickly switched, meets the requirements of different departments.
-
Abundant analysis functions: 23 arrhythmia, full-lead ST-segment and pacing analysis; multiple calculation functions (such as calculation of drug concentration, titration table, hemodynamics, ventilation, oxygenation, renal function, ).
-
With the functions of waveform freezing, holographic waveform storage and review.
-
Automatic data storage in case of power failure.
-
Built-in large-capacity detachable lithium battery, ensures uninterrupted monitoring.
-
Fanless design, ensures no noise and low power consumption during working.
-
Connect to Central Monitoring System developed by our company by WiFi/Wired, convenient to transmit patient's real-time information by mode, be applied to clinical monitoring, avoids affecting patient's rest, one-button switching, reduces the workload of medical staff.
Performance:
ECG
3-lead: I, II, II
5-lead: l, II, Il, aVR, aVL, aVF, V
12-lead: l, ll, Il, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
-
Gain: 5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
-
Scanspeed: 25mm/s, 12.5mm/s, 25mm/s, 50mm/s
-
HR
Measurement and alarm range:
Adult: 15~300 bpm
Pediatric/neonate: 15~350 bpm
Resolution: 1 bpm
Accuracy: ±1% or ±1bpm, whicheveris greater.
Measurement and alarm range: -2.0mV~+2.0 mV
Accuracy: -0.8mV~+0.8mV: ±0.04mV or ±10%, whicheveris greater.
Other range: unspecified
-
Arrhythmiaanalysis: 23 kinds
Pacemaker: yes
RESP
-
Measurementmode: impedance
-
Respirationrate:
Measurement range: 0 rpm~150 rpm
Resolution: 1 rpm
Accuracy: 7~150 rpm, ±2rpm or±2%, whicheveris greater;
0~6 rpm, unspecified
-
Apneaalarm: 10~60s.
-
Scanspeed: 25mm/s, 12.5 mm/s, 25mm/s, 50mm/s.
NIBP
-
Measurementmode: Oscillometry
-
Workingmode: Manual/AUTO/Continuous/Order
-
Measurementinterval in AUTO mode: 1 min, 2 min, 5 min, 3 min,
5 min, 10 min, 15 min, 20min, 30min, 1h, 1.5h, 2h, 3h, 4h, 8h
-
Measurement period in Continuous mode: 5 minutes
-
Staticpressure Measurement range: 0mmHg~300 mmHg (0.0kPa~40kPa)
-
Resolution: 1mmHg
-
Cuffpressure accuracy: ±3mmHg
-
Measurementaccuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Adult mode: 297±3mmHg
Pediatric mode: 260±3 mmHg
Neonatal mode: 147±3mmHg
SpO₂
-
Measurementrange: 0%~100 %
-
Resolution: 1%
-
Accuracy: 70%~100%, ±2%;0%~69%, unspecified
-
PR
Measurement and alarm range: 25~250 bpm
Accuracy: ±2 bpm or±2%, whicheveris greater.
TEMP
-
Channel: dual-channel
-
Measurementrange: 0~50℃
-
Resolution: 1℃
-
Accuracy: ±0.1℃
CO₂
-
Measurement mode: mainstream or sidestream
-
CO₂ measurementrange: 0~150 mmHg
-
Accuracy:
0~40mmHg: ±2mmHg;
41~70mmHg: ±5%;
71~100mmHg: ±8%;
101~150mmHg: ±10%
Sidestream: 2~120 rpm;
Mainstream: 2~150 rpm.
-
AwRRaccuracy: ±1rpm
-
Apneaalarm: yes
IBP
-
Channel: 2-channel(Expandable 4-channel)
-
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
-
Measurement and alarm range: -50~300 mmHg.
-
Resolution: 1mmHg.
-
Accuracy: ±2%or 1 mmHg, whicheveris greater.
Power supply: 100V-240V~50Hz/60 Hz.
Safety class: classl, defibrillation-proof type CF applied part.
Waterproof degree: IPX2.
Accessories
Category Accessory name
ECG lead cable 5-lead, American standard
ECG electrode adult, one packet(20 pcs)
SpO2 probe (integrated) Digital fingertip SpO2 probe for adult (CMS-N-SPO2 6P, 3M, yellow)
TEMP probe R25=2.252K TEMP probe, skin surface type, CMS-N-TEMP 2P, PVC, L=3M
NIBP cuff BP extension tube, L=3M
Adult cuff (25 ~ 35 cm)
Physical characteristic
Component Dimensions (l×w× h) Weight Remark
Host 351mm×234mm×60mm About 2.1kg Standard configuration (including battery), weight does not include recorder and accessories;
Optional configuration
No. Function name
1 Recorder module
2 CO2 module
3 IBP module
4 12-lead ECG
5 Central Monitoring System