Function
Parameters: ECG, RESP, SpO2, PR, NIBP, TEMP, CO2 (optional)
1)ECG Heart rate (HR)
ECG waveform
2)RESP Respiration rate (RR)
Respiration waveform
3)SpO2 Oxygen saturation (SpO2)
Plethysmogram (PLETH) waveform
Pulse rate (PR)
Bar graph
4)NIBP Systolic pressure (SYS), Diastolic pressure (DIA), Mean pressure (MEAN)
5)TEMP
6)CO2 (optional) EtCO2, INSCO2, AWRR
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation.
Performance
▲ECG
Lead mode: 3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
Waveform: 3-lead, 1 channel
5-lead, 2-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
Resolution: 1 bpm
▲RESP
Method: R-F (RA-LL) impedance
Respiration rate:
Measurement and alarm range: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy:
7 rpm ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 rpm ~ 6 rpm, unspecified
Apnea alarm: 10 ~ 40 s
Scan speed: 12.5 mm/s, 25 mm/s
▲NIBP
Measurement mode: Oscillometry
Working mode: Manual / AUTO / STAT
Measurement interval in AUTO mode: 1 / 2 /3 / 4 / 5/ 10 / 15 / 30 / 60 / 90 / 120 / 180 / 240 / 480 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 0 ~ 300 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ± 3 mmHg
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult: 297±3 mmHg
Pediatric: 260±3 mmHg
Neonate: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1 %
Accuracy: 70 % ~ 100 %: ± 2 %;
0 % ~ 69 %: unspecified
▲PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
▲TEMP
Measurement and alarm range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Accuracy: ± 0.1 ℃
▲CO2(optional)
Measurement mode: mainstream or sidestream
Measurement range: 0 ~ 150 mmHg
Resolution:
0.1 mmHg: 0 ~ 69 mmHg
0.25 mmHg: 70 ~ 150 mmHg
Accuracy:
0 ~ 40 mmHg: ±2 mmHg
41 ~ 70 mmHg: ±5 %
71 ~ 100 mmHg: ±8 %
101 ~ 150 mmHg: ±10 %
AwRR measurement range:
Sidestream: 2 ~ 120 rpm
Mainstream: 2 ~ 150 rpm
Accuracy: ±1 rpm
Apnea alarm: yes
▲Power supply: 100 V-240 V ~ 50 Hz/60 Hz
▲Safety class: Class II
Accessories
1)VET SpO2 probe
1)VET NIBP cuff
2)NIBP extension tube
2)ECG lead cable
3)ECG electrode clips
4)TEMP probe
3)Power cord
4)User Manual
Physical characteristic
Dimension: 168 mm (L) × 56 mm (W) × 98 mm (H)
Weight: <1.0 Kg (excluding accessories)