▲ECG
Lead mode: 3-lead:I, II, III
5-lead:I, II, III, aVR, aVL, aVF, V
Waveform: 3-lead, 1-channel
5-lead, 2-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV, 40 mm/mV
Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350 bpm
Resolution: 1 bpm
Accuracy: ±1 % or ±1 bpm, whichever is greater
Alarm accuracy: ± 2 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0 mV~+2.0 mV
Accuracy: -0.8 mV~+0.8 mV:±0.04 mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON
T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
▲RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: 7 rpm~150 rpm, ±2 rpm or ±2%, whichever is greater; 0 rpm~6 rpm, no definition;
Apnea alarm: 10~40 s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
▲NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10~270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297 mmHg±3 mmHg
Pediatric mode: 240 mmHg±3 mmHg
Neonatal mode: 147 mmHg±3 mmHg
▲SpO2
Measurement and alarm range: 0~100 %
Resolution: 1 %
Accuracy: 70~100 %: ±2 %
0~69 %: unspecified
▲PR
Measurement and alarm range:25~250 bpm
Measurement accuracy: ±2 bpm or ±2 %, whichever is greater
▲TEMP
Channel: dual-channel
Measurement and alarm range: 0~50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
▲EtCO2
Method: Sidestream or Mainstream
Measuring Range: 0~150 mmHg
Resolution:
0~69 mmHg: 0.1 mmHg
70~150 mmHg: 0.25 mmHg
Accuracy:
0~40 mm Hg: ±2 mm Hg
41~70 mm Hg: ±5 %
71~100 mm Hg: ±8 %
101~150 mm Hg: ±10 %
AwRR Range: 2~150 rpm
AwRR Accuracy: ±1 rpm
Apnea Alarm: Yes
▲IBP
Channel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range: -50~300 mmHg
Resolution: 1 mmHg
Accuracy: ±2% or 1mmHg, whichever is greater
▲Power supply: 100 V-240 V ~ 50 HZ/60 Hz
▲Safety classification: Class I, type CF defibrillation-proof applied part
▲Harmful liquid proof degree: IPX2