CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor 2-IBP+Printer - CONTEC
  • SKU: 3.05.014.05.026
  • Availability: in stock Many in stock Out of stock You can purchase this product but it's out of stock

CONTEC CMS9000 Capnograph CO2 monitor Vital Signs ICU/CCU Patient Monitor IBP+Printer 12.1'' TFT color LCD

$898.00 $999.00
  • Barcode: 658126913409
DESCRIPTION

Introduction 

This device can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and dual-channel TEMP, etc. It integrates parameter measurement module, display and recorder in one device to form a compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for patient moving.

Functions

Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP, ETCO2, IBP

1)ECG   Heart rate(HR)

      ECG waveform 

      Arrhythmia and ST-segment analysis

2)RESP   Respiration rate(RR)

      Respiration waveform

3)SpO2   Pulse oxygen saturation(SpO2)

      Plethysmogram(PLETH) waveform

      Pulse rate(PR)

      Bar graph

4)NIBP   Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)

5)TEMP   T1, T2, TD

It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.

Features

1)12.1'' TFT color LCD, 15-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).

2)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.

3)All-round monitor for adult, pediatric and neonate.

4)With standard interface, OxyCRG, trend chard, big character interface and view bed, convenient to observe.

5)Waveform, parameter color and location can be set optionally.

6)Operation mode: keys and buttons.

7)Display up to 8-channel waveform on one screen.

8)Display 7-lead ECG waveform on one screen, and with ECG step function.

9)Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.

10)HRV analysis function.

11)NIBP measurement mode: Manual/AUTO/Continuous, storage for 4800-group NIBP data.

12)Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.

13)Drug concentration calculation and titration table functions. 

14)Print trend table data with one-key.

15)Connect to Central Monitoring System by 3G, WiFi or wired mode.

16)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.

17)Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).

Performance 

ECG 

Lead mode: 3-lead or 5-lead

Lead selection: I, II, III, aVR, aVL, aVF, V

Waveform: 5-lead, 2-channel

3-lead, 1-channel

Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV

Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s

HR: 

Measurement and alarm range: 15~350bpm

Accuracy: ±1% or ±1bpm, whichever is greater

Alarm accuracy: ± 2 bpm

Resolution: 1 bpm

ST-segment monitoring:

Measurement and alarm range: -2.0mV~+2.0mV

Accuracy: -0.8mv~+0.8mv  ±0.04mv or ±10%, whichever is greater 

Other range: unspecified 

Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

Pacemaker: yes

RESPIRATION

Method: R-F(RA-LL) Impedance 

Respiration rate: 

Measurement and alarm range: 0~150rpm

Resolution: 1 rpm

Measurement accuracy: ±2 rpm

Alarm accuracy: ±3 rpm

Apnea alarm: 10~40s

Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s

NIBP

Method: Oscillometry

Mode: Manual/AUTO/Continuous 

Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes

Measurement period in Continuous mode: 5 minutes

Measurement and alarm range: 10 ~ 270 mmHg

Resolution: 1 mmHg

Cuff pressure accuracy: ±3 mmHg

Measurement accuracy:

Maximal mean deviation: ±5 mmHg

Maximal standard deviation: 8mmHg

Over-pressure protection:

Adult mode: 297±3 mmHg

 Pediatric mode: 240±3 mmHg

 Neonatal mode: 147±3 mmHg

SpO2

Measurement and alarm range: 0 ~ 100%

Resolution: 1%

Measurement accuracy: 70%~100%: ±2%; 

0%~69%: unspecified 

PR

Measurement and alarm range: 30 ~ 250 bpm

Measurement accuracy: ±2 bpm or ±2%, whichever is greater

TEMP

Channel: dual-channel

Measurement and alarm range: 0 ~ 50℃

Resolution: 0.1 ℃

Accuracy: ±0.1 ℃

EtCO2

Method: Sidestream or Mainstream 

Measurement range for CO2: 0~150mmHg

Resolution for CO2:

 0 ~ 69 mmHg: 0.1 mmHg

 70 ~ 150 mmHg: 0.25 mmHg

Accuracy for CO2:

 0 ~ 40 mmHg: ±2 mmHg

 41 ~ 70 mmHg: ±5%

 71 ~ 100 mmHg: ±8%

 101 ~ 150 mmHg: ±10%

AwRR range: 2~150 rpm

AwRR accuracy: ±1 rpm

Apnea alarm: yes

IBP

Channel: dual-channel

Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

Measurement and alarm range: -10~300 mmHg

Resolution: 1 mmHg

Accuracy: ±2% or 1mmHg, whichever is greater

Power supply:AC 100-240V, 50/60Hz

Safety classification: Class I, type CF defibrillation-proof applied part

Accessories

Adult fingertip SpO2 probe (5-pin)

Adult NIBP cuff 

NIBP extension tube

ECG lead cable

ECG electrode

Temperature probe

Power cord 

User Manual (GB only)on request for other Language

CO2 module

IBP module

Print paper

Physical characteristic

Dimension: 319 mm(L) × 161 mm(W) × 269 mm(H)

Weight: 3.6 Kg

ADDITIONAL INFORMATION
Color: Blue, Purple, White
Size: 20, 24
Material: 100% Polyester
REVIEWS
GENERAL TAB

RELATED PRODUCTS

BACK TO TOP